Specifications |
Temperature Range
and Resolution: |
|
-200 ° C to +400 C, 1 ° C resolution
- l00°C to + 199.9°C, 0.1°C resolution |
Differential Temperature
Range: |
|
- 19.99° to + 19.99°. Linearization
centered at 40°C. 0.01°C resolution. |
Accuracy, 1° range
0.1° range
diff. range |
|
1°C + 1 least significant digit
0.1°C + 1 least significant digit
0.01°C + 1 least significant digit |
| Repeatability: |
|
+ one least significant digit |
| Calibration: |
|
Conforms to National Institute of Standards and Technology tables (Monograph 125) |
| Sensors: |
|
Any Physitemp type T thermocouple |
Ambient Operating
Range: |
|
15 - 45°C |
| Readout: |
|
Liquid crystal, 3 1/2 digits |
| Batteries: |
|
BAT-10 -- 4 alkaline "C" cells
(battery life, 1000 hours)
BAT-10R (rechargeable unit) - 4 Ni-Cads |
| Analog Output: |
|
Non-linearized set at 1.6V corresponding to temperature of 401°C |